Medi-Stim Neurocare TENS TENS Unit

Medi-Stim Neurocare TENS TENS Unit

$45.99

Features:

  • Dual channel (4-lead) digital TENS
  • Five treatment modes (Normal, Burst, Modulated Rate, Modulated Width, Strength Duration)
  • Attractive high-tech design with metallic finish w/ LCD display
  • Hinged protective cap for intensity dials
  • Easy-to-use push-button controls
  • Locking battery/control panel
  • Durable metal wire belt clip
  • Treatment timer 5-60 minutes, or Continuous
  • Compliance monitor records up to 999 hours and 59 minutes of use
  • 3-year warranty

Comes with 1 pkg of 4 reusable self-adhering electrodes, 1 set 43” pin lead wires, 1 each 9V heavy duty battery, carrying case, and instruction manual

Leadwire reorder # LW46SPLE

Description

Technical Specifications:

  • Channel: Dual, isolated between channels
  • Amplitude: Adjustable, 0 – 90 mA; max output 90 mA (peak to peak) into 500 ohm each channel
  • Output Voltage: Adjustable, 0 – 45 V; max output 45V (peak to peak) into 500 ohm each channel
  • Waveform: Asymmetrical biphasic square pulse
  • Pulse Rate: Adjustable, 1 ~ 150 Hz, 1 Hz/step
  • Pulse Width: Adjustable, 30 ~ 300 µS, 10 µS/step

Modes Five (Normal, Burst, Modulated Rate, Modulated Width, Strength Duration)

  • Normal (N) Continuous output, pulse rate and pulse width are adjustable.
  • Burst Mode (B) Burst rate adjustable, 0.5 ~ 5 Hz; Pulse width adjustable, 30 ~ 300 µS, Frequency fixed = 100 Hz
  • Modulation (MR) Pulse rate is automatically varied in a cyclic pattern over an interval of nominally 10 seconds (in max 150 Hz). Pulse rate decreases linearly over a period of 5 seconds from the control setting value to a value which is 70% less, then increases linearly over a 5 seconds period to its original value. The cycle is then repeated.
  • Modulated Width (MW) Pulse width is automatically varied in a cyclic pattern over an interval of nominally 10 seconds (in max 300 µS). Pulse width decreases linearly over a period of 5 seconds from the control setting value to a value which is 70% less, then increases linearly over a 5 seconds period to its original value. The cycle is then repeated.
  • Strength Duration (SD) Pulse rate and pulse width are varied in a cyclic pattern over an interval of nominally 10 seconds. so the pulse width is always increasing while the pulse rate is decreasing and vice-versa. Pulse rate decreases linearly over a period of 5 seconds from the control setting value to a value which is 90% less (width equal to setting). Pulse width decreases linearly over a period of 3 seconds from the control setting value to a value which is 38% less (rate equal to setting). The cycle is then repeated. Pulse rate (from 1 ~ 150 Hz) and pulse width (from 30 ~ 300 µS) are fully adjustable.

Timer: Adjustable, 5, 10, 20, 30, 40, 50, 60 minutes or Continuous
Power Supply: One 9-volt battery
Size: 10 x 6 x 2.1 cm
Weight: 120 grams with battery